| Model Number |
8888561050 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
An investigation is currently underway.Upon completion, the results will be forwarded.
|
| |
|
Event or Problem Description
|
|
The customer reported the blue handle of the trocar has broken off three times in a row.No patient was harmed.
|
| |
|
Additional Manufacturer Narrative
|
|
The device history record (dhr) review showed that all acceptance criteria inspections per established sampling levels were within acceptable limits during the production process.No sample was available for evaluation.Without a sample we were unable to perform a thorough investigation to determine the root cause or implement any corrective action.We were unable to confirm the reported condition.No root cause can be determined.A corrective and reported action (cap)a will not be initiated.This complaint will be used for tracking and trending purposes.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: h6 type of investigation code was added.
|
| |
|