| Model Number |
CX*FX25E |
| Medical Device Problem Code |
Device Damaged Prior to Use (2284)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/06/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
D6a: implanted date: device was not implanted.D6b: explanted date: device was not explanted.E3: occupation: ccp.The actual device has not been returned for evaluation.The investigation is currently ongoing.A follow-up report will be submitted once the investigation is complete.The manufacturing record and the shipping inspection record of the actual product found no anomaly.No other similar report of the product with the involved product code/lot was found.Terumo medical products (tmp) (importer) registration no.2243441 is submitting this report on behalf of ashitaka factory of terumo corporation (manufacturer) registration no.9681834.
|
| |
|
Event or Problem Description
|
|
The user facility reported that the caipox device involved was damaged.The event occurred out of the box.The event did not result in a delay in the surgical procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted as follow up no.1 to provide the device return date in section d9, update section h3, and to provide the completed investigation results.Visual and magnifying inspections of the actual sample found a crack was found in the connecting part between the female connector and tubing on the sampling line connected to the oxygenator blood outlet port.The surface of female connector had some marks, and the female connector was also found to have cracks.Leak test of the actual sample found the colored saline solution was flowed into the actual sample by a head.As a result, leakage was found from the crack.X-ray fluoroscopic inspection of the actual female connector found that the tube inside the female connector was also found to have cracks.Simulation test: an impact load was applied to the female connector of current product.As a result, it was found that a crack similar to the actual sample appeared.X-ray fluoroscopic inspection of the cracked sample was performed.It was found that the crack similar to the actual sample was simulated on the tube inside the female connector.The manufacturing record and the shipping inspection record of the actual sample found no anomaly.No other similar report of the product with the involved product code/lot number was found.Based on the investigation result, it was confirmed that a crack was found in the connecting part between the female connector and tubing on the sampling line.As a possible cause of the crack on the tube, it was inferred that some impact load was applied to the female connector that exceeding the limit strength of product.However, from the condition of actual sample, it was not possible to clarify exactly when the crack occurred.The instructions for use (ifu) (capiox fx25) has the following cautions: "do not use if the package or device is damaged (e.G., cracked) or any of the port caps are off." "if the product is dropped during set-up, do not use it.Replace with another device.".
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted as follow up no.2 to provide a correction to sections d1: brand name, d4: model number, d4: catalog number, d4: lot number, d4: expiration date, d4: udi, and h4: device manufacture date.The udi was updated to match the correct format.Due to internal review, a correction was made to section d3.
|
| |
|