| Catalog Number |
42527000510 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/30/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that two tibial articular surface provisional instruments fractured.There was no patient impact and no adverse events have been reported as a result of the malfunction.It was reported that no further information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).D10: medical product: tibial articular surface provisional right size ef: catalog#42527000505, lot#64312234; multiple mdr reports have been filed for this event.Please see associated report: 0001822565-2023-03703.The product has been received by zimmer biomet and the investigation is in process.Upon completion of the investigation, a follow-up mdr will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.H6: proposed component (annex g) code is mechanical (g04) - provisional top.A visual inspection of the returned item found it to exhibit signs of repeated use and is fractured.Device history record was reviewed and no discrepancies related to the reported event were found.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further event information at time of this report.
|
| |
|