Continuation of d10: product id 97745ac lot# serial# (b)(6).Product type accessory.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Information was received from a patient who was implanted with an implantable neurostimulator (ins) for non-malignant pain.It was r eported that the patient can't charge up controller for "about the last week and a half or two weeks ago." resetting controller didn't fix issue.Controller wont power on with battery pack out and ac power plugged in, and the green light is on, on ac power plug.During the call they put in aa batteries and screen does come on.Controller port has a few pins in the center that look blackened out.The issue was not resolved.An email was sent to the repair department to replace the controller.Patient called back and stated they received the replacement controller, but the controller is still not powering on with their original battery pack or the ac power supply connected.Patient stated their replacement controller powers on with aa batteries inserted.Agent had patient remove the batteries from the replacement controller and connect to the ac power supply; the controller did not power on.Agent sent an email to repair to replace the ac power supply.
|