• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION HD CARTRIDGE LINES; SET, TUBING, BLOOD, WITH AND WITHOUT ANTI-REGURGITATION VALVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION HD CARTRIDGE LINES; SET, TUBING, BLOOD, WITH AND WITHOUT ANTI-REGURGITATION VALVE Back to Search Results
Catalog Number 101025
Medical Device Problem Codes Crack (1135); Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/25/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that there was a crack in the cassette of a gambro cartridge which resulted in an external saline leak.The leak was discovered while priming the set prior to patient use.There was no patient involvement.No additional information is available.
 
Additional Manufacturer Narrative
E1: initial reporter address: (b)(6).E1: initial reporter phone no: (b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Additional Manufacturer Narrative
The actual device was not available; however, a photograph of the sample was provided for evaluation.During visual inspection of the provided photographic samples, a crack was observed in the arterial and venous chamber of cartridge.The crack was observed inside of the chamber to the atmosphere, suggesting that the crack occurred inside to outside.The reported condition was verified.A total of 8 retention samples were visually and functionally tested but no deviations were found.The retention samples were found within the specifications.Due to the nature of the provided samples, no further testing could be performed; therefore, the cause of the condition could not be determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HD CARTRIDGE LINES
Common Device Name
SET, TUBING, BLOOD, WITH AND WITHOUT ANTI-REGURGITATION VALVE
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - TIJUANA BAJA
blvd pacifico 10014
parque industrial pacifico
tijuana baja california CP 22 643
MX   CP 22643
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key18400311
Report Number8030638-2023-00031
Device Sequence Number4196689
Product Code FJK
UDI-Device Identifier07332414007836
UDI-Public(01)07332414007836
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
K070414
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/15/2026
Device Catalogue Number101025
Device Lot NumberMA0123032074
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/30/2023
Supplement Date Received by Manufacturer01/31/2024
Initial Report FDA Received Date12/27/2023
Supplement Report FDA Received Date02/05/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/09/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-