| Catalog Number |
MS18597 |
| Medical Device Problem Codes |
No Device Output (1435); Failure to Sense (1559)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/15/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is ongoing; results will be provided in a follow-up report.H3 other text : on-going.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient monitor lost the co2 trace during an anesthetic procedure and was alarming for battery error.The delta patient monitor was connected via an infinity docking station (ids) to a dräger fabius anesthesia workstation; the gas readings were provided by a dräger scio anesthetic gas monitor.It was mentioned in the report that the event did not lead to consequences for the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
A draeger technician went on site and localized the reported issues to a failed power supply which was replaced.No further issues have been reported.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient monitor lost the co2 trace during an anesthetic procedure and was alarming for battery error.The delta patient monitor was connected via an infinity docking station (ids) to a draeger fabius anesthesia workstation; the gas readings were provided by a draeger scio anesthetic gas monitor.It was mentioned in the report that the event did not lead to consequences for the patient.
|
| |
|