• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR SUPERA SELF-EXPANDING STENT SYSTEM; SELF EXPANDING PERIPHERAL STENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR SUPERA SELF-EXPANDING STENT SYSTEM; SELF EXPANDING PERIPHERAL STENT SYSTEM Back to Search Results
Catalog Number S-45-060-120-P6
Medical Device Problem Codes Material Separation (1562); Improper or Incorrect Procedure or Method (2017); Failure to Advance (2524); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/09/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Manufacturer's investigation is still pending at this time.Results and conclusions will be provided in the final report.Medwatch report # mw5148102.
 
Event or Problem Description
User facility medwatch report (mw5148102) received stating: abbott supera stent delivery device was inserted over the wire and was unable to cross the lesion.The device was removed.Cardiologist noticed a piece of the delivery system missing and the stent was off the delivery device.The missing piece was snared and retrieved.Additional information received from the site: the vessel treated was the distal left popliteal artery that is severely diseased at 90-95%.There was resistance during advancement and failed to cross due to the anatomy.Nothing remains in the patient.There was no clinically significant delay in the procedure.
 
Additional Manufacturer Narrative
Medical device problem code 2017 - against resistance the device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents from this lot.Reportedly, force was applied in attempts to advance the device.It should be noted that the supera peripheral stent system instructions for use states: precaution: should unusual resistance be felt at any time during stent system advancement or stent deployment the entire system should be removed together with the introducer sheath or guiding catheter as a single unit.The investigation determined the reported difficulties appear to be related to circumstances of the procedure as it is likely that during advancement resistance was met with the 90-95% severely diseased anatomy resulting in the reported failure to advance.Interaction/manipulation of the device using force and/or inadvertent mishandling ultimately resulted in the reported material separation and the reported device dislodged or dislocated.The treatment appears to be related to the operational context of the procedure as reportedly the missing piece was snared and retrieved.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUPERA SELF-EXPANDING STENT SYSTEM
Common Device Name
SELF EXPANDING PERIPHERAL STENT SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 2024168
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key18401390
Report Number2024168-2023-14493
Device Sequence Number4020594
Product Code NIP
UDI-Device Identifier08717648225949
UDI-Public08717648225949
Combination Product (Y/N)N
Initial Reporter StateWV
Initial Reporter CountryUS
PMA/510(K) Number
P120020
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/24/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2024
Device Catalogue NumberS-45-060-120-P6
Device Lot Number2101962
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/06/2023
Supplement Date Received by Manufacturer01/05/2024
Initial Report FDA Received Date12/27/2023
Supplement Report FDA Received Date01/24/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/19/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age83 YR
Patient SexFemale
Patient Weight59 KG
-
-