• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 ATTUNE SHIM SZ7 8MM; TIBIAL INSERT TRIALS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 ATTUNE SHIM SZ7 8MM; TIBIAL INSERT TRIALS Back to Search Results
Catalog Number 254500673
Medical Device Problem Code Scratched Material (3020)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/13/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : according to the information received, "[all items have gouges except for the impactor which is cracked]" the product was not returned to depuy synthes, however photos were provided for review.See attachment (pcs 13.12.2023).The photo investigation revealed that the device 254500673, attune shim sz7 8mm was found scratched and chipped on the surface.This type of damage is consistent with the device being in contact with other tools like; forceps, tweezers or other tools with and edge.Properly handling and attention to the approved use of the device diminishes the risk of failure.Since the device was not returned, a dimensional inspection cannot be performed.The overall complaint was confirmed as the observed condition of the device 254500673, attune shim sz7 8mm would contribute to the complained device issue.Based on the investigation findings, potential cause can be attributed to unintended use error and it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot
=
> the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.H6: component code: appropriate term/code not available (g07002) used to capture no findings available.This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported that the attune shim was cracked.The product damage documented in this pc has been identified during loan kit inspection, by the loan kit technician.The event date and alert date are the date that the inspection took place.There is no surgeon, procedure, or patient details available.No further information can be obtained as the case was not reported by the customer.
 
Additional Manufacturer Narrative
Product complaint (b)(4).Investigation summary: all items have gouges, except for the impactor.Which is cracked.The product damage documented in this pc has been identified, during loan kit inspection, by the loan kit technician.The event date and alert date are the date that the inspection took place.There is no surgeon, procedure or patient details available.No further information can be obtained, as the case was not reported by the customer.The device associated with this report was returned to depuy synthes for evaluation.Visual analysis of the returned device found, signs of chipped of and some scratches at the surface of the sample.The reported allegation can be confirmed.This type of damage is consistent with the device being in contact with other tools like: forceps, tweezers or other tools with and edge.Properly handling and attention to the approved use of the device diminishes the risk of failure.A dimensional inspection was not performed, since it was not applicable to the complaint condition.A functional test was not performed, since it was not applicable to the complaint condition.The overall complaint was confirmed.As the observed, condition of the device attune shim would contribute to the complained device issue.Based on the investigation findings, the potential cause is traced to unintended use error.And it has been determined, that no corrective and/or preventative action is required.As part of depuy synthes quality process, all devices are manufactured, inspected and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information that has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: added: d9 if information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Corrected: h3.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE SHIM SZ7 8MM
Common Device Name
TIBIAL INSERT TRIALS
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key18404564
Report Number1818910-2023-26076
Device Sequence Number3968631
Product Code HWT
UDI-Device Identifier10603295133551
UDI-Public10603295133551
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number254500673
Device Lot NumberMVMDXT350
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/19/2023
Supplement Date Received by Manufacturer02/16/2024
03/05/2024
Initial Report FDA Received Date12/27/2023
Supplement Report FDA Received Date02/21/2024
03/05/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/02/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-