| Model Number |
DDBC3D4 |
| Medical Device Problem Code |
Device Sensing Problem (2917)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/06/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient heard an audible alert.A remote monitoring transmission indicated that the right ventricular (rv) lead triggered a lead integrity alert (lia) due to high rate non-sustained episodes and sensing integrity counter (sic).Additionally, the rv lead displayed oversensing of noise.The right atrial (ra) lead triggered an alert for high atrial thresholds, which were also noted to be variable.Further, the ra lead displayed small p waves and atrial undersensing with a subsequent lack of detection and termination of atrial tachycardia/atrial fibrillation (at/af) and inappropriate atrial pacing.The implantable cardioverter defibrillator (icd), ra lead, and rv lead remain in use. no patient complications have been reported as a result of this event.
|
| |
|
Search Alerts/Recalls
|