| Catalog Number |
306616 |
| Medical Device Problem Code |
Complete Blockage (1094)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/04/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
B.3.The date received by manufacturer has been used for this field.H.3.If a device evaluation and/or device history review is completed, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that the bd safetyglide needle was clogged/blocked.The following information was provided by the initial reporter: "no injuries reported.Customer states some of the needles do not have any holes.They are solid." product#: 306616.Lot#: 2202914.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4) ¿ follow up mdr for device evaluation: no samples or photos received by our quality team for evaluation.No root cause can be determined as no samples were received.Furthermore, a device history record review showed no rejected inspections or quality issues during the production of the provided lot number that could have contributed to the reported defect.During review of the process, process variations during the needle lubricant application can create clogged needles.Corrective and preventative actions have been implemented.Several quality initiatives have been implemented on our manufacturing line to ensure that the needle lubricant application is properly applied during the manufacturing process.Additionally, a monitoring program is also in place to verify the needle lubricant is applied uniformly to the needle.We appreciate you taking the time to bring this observation to our attention.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored monthly.Our business team regularly reviews the collected data for identification of emerging trends.H3 other text : see h10 manufacture narrative.
|
| |
|
Event or Problem Description
|
|
No additional information.
|
| |
|