• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND S5 MAST ROLLER PUMP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND S5 MAST ROLLER PUMP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 10-88-00
Medical Device Problem Code Pumping Stopped (1503)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/01/2023
Type of Reportable Event Malfunction
Event or Problem Description
Livanova deutschland received a report that a s5 mast roller pump stopped during the case and there was no alarm, either audible or visual.Customer tried to restart the pump without success, therefore hand crank was started; normal alarms of runaway motor due to hand cranking appeared.The pump was then turned off and on again and stopped hand cranking: the value on the screen went up, but the pump didn't turn.Customer believed to have the pump lid overridden at this point, however no alarm was working on the involved arterial pump.Lastly, the affected pump was changed with a centrifugal one and everything started again.Surgery was completed with no patient injury.
 
Additional Manufacturer Narrative
A.1.-a.5.There was no patient involvement.H10: livanova deutschland manufactures the s5 mast roller pump.The incident occurred in (b)(6).Through follow-up communication livanova learned that the involved arterial pump stopped during case on (b)(6) 2023 at around 17:05 or 17:06.A livanova field service representative was dispatched to the facility to investigate the device and could not reproduce the reported issue: affected pump was working properly when turned on.Full functional testing was completed on the pump, also with the sensors connected, no issue arisen.Moreover, a completed a simulation of what occurred was performed and when hand cranking was done, the alarms still worked.Pump was then changed out for a loan and has been taken back to the workshop for review.Loan roller pump was installed and fully tested, also with sensors, without issues.In addition, e/p (electronics and power) pack was changed as per customer's request.The serial read-out of the affected pump (real time device parameters and setting recording file) was analyzed and revealed that on date of event the following error message was stored: 01.12.23!17:06:47.443 15=canap 50ah=control_stop 8 2, that suggests that the pump was wrongly linked to the cardioplegia module.Livanova is investigating if this is the root cause of the pump stop.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Additional Manufacturer Narrative
Based on the log files analysis and considering that the pump stopped without any alarm, it can be assumed that the arterial pump was wrongly linked to the cardioplegia module, consequently, it is highly probable that the cardioplegia bubble alarm stopped the arterial pump.The risk is in the acceptable region.No specific action was currently deemed necessary.Livanova will keep monitoring the market.
 
Event or Problem Description
See initial report.
 
Event or Problem Description
See initial report.
 
Additional Manufacturer Narrative
Udi related data quality updates only d2a.Common device name has been corrected in the dedicated section.D2b.Device product code has been corrected in the dedicated section.G.4 pma/510(k) number has been corrected in the dedicated section.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 MAST ROLLER PUMP
Common Device Name
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND GMBH
lindberghstrasse 25
munich 80939
GM   80939
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key18407711
Report Number9611109-2023-00661
Device Sequence Number7201465
Product Code DTQ
UDI-Device Identifier04033817900474
UDI-Public(01)04033817900474(11)181212
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K103762
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/16/2024
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number10-88-00
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/02/2023
Supplement Date Received by Manufacturer02/29/2024
08/08/2024
Initial Report FDA Received Date12/28/2023
Supplement Report FDA Received Date03/27/2024
08/21/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/12/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-