• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AVANOS MEDICAL INC. ON-Q PAIN RELIEF SYSTEM WITH SELECT-A-FLOW, 600 ML, 2-14 ML/HR; ELASTOMERIC - SAF

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AVANOS MEDICAL INC. ON-Q PAIN RELIEF SYSTEM WITH SELECT-A-FLOW, 600 ML, 2-14 ML/HR; ELASTOMERIC - SAF Back to Search Results
Model Number CB6004
Medical Device Problem Code Improper Flow or Infusion (2954)
Health Effect - Clinical Code Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Type of Reportable Event Malfunction
Event or Problem Description
Avanos medical inc.Received a single report that referenced three incidences, which were associated with separate units, involving three different patients.This is the first of three reports.Refer to 2026095-2023-00114 for the second report.Refer to 2026095-2023-00115 for the third report.Fill volume: 600 ml.Flow rate: unknown.Procedure: unknown.Cathplace: unknown.Infusion start time: unknown.Infusion stop time: unknown.It was reported, the pump emptied at least a day prior to what was expected.One patient experienced symptoms of local anesthetic systemic toxicity (last).No medical interventions reported.
 
Additional Manufacturer Narrative
H6: health effect - clinical code: appropriate term / code not available: local anesthetic systemic toxicity (last).The actual complaint product was not returned for evaluation.A review of the device history record is not possible as no lot number was provided.A root cause can not be determined.All information reasonably known as of 26 dec 2023 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the reported event but is relaying the information provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ON-Q PAIN RELIEF SYSTEM WITH SELECT-A-FLOW, 600 ML, 2-14 ML/HR
Common Device Name
ELASTOMERIC - SAF
Manufacturer (Section D)
AVANOS MEDICAL INC.
5405 windward parkway
alpharetta GA 30004
Manufacturer (Section G)
AVENT S. DE R.L. DE C.V.
ave noruega edificio d-1b
fraccionamiento rubio
tijuana b.c. 22116
MX   22116
Manufacturer Contact
lisa clark
5405 windward parkway
alpharetta, GA 30004
4704485444
MDR Report Key18408670
Report Number2026095-2023-00113
Device Sequence Number8723472
Product Code MEB
UDI-Device Identifier00193494134761
UDI-Public00193494134761
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K063530
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Distributor
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 12/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberCB6004
Device Catalogue NumberN/A
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/06/2023
Initial Report FDA Received Date12/28/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-