Model Number EC300
Medical Device Problem Code Power Problem (3010)
Health Effect - Clinical Code Unspecified Tissue Injury (4559)
Date of Event 11/19/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Medical safety has reviewed the alleged event that during a craniotomy, the powered system "was insufficient and the bone flap could not be removed.Therefore, it was replaced with a hand drill for the craniotomy, which resulted in a larger craniotomy bone seam and greater trauma to the patient." this case and report of "serious harm" and the relationship to the device (ipc console ec300) is unassessable due to incomplete information regarding the manner of alleged device failure, and the reported injury/impact to the patient.Medtronic's attempts to obtain additional information did not add information that would allow for more clarity of the product performance and patient status.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that during craniotomy surgery, when the comprehensive surgical power system was used, the power was insufficient and the bone flap could not be removed.Therefore, it was replaced with a hand drill for the craniotomy, which resulted in a larger craniotomy bone seam and greater trauma to the patient.Additional information received that the power was insufficient throughout the entire surgery.The bur was moving, the drilling speed was not as expected, and it couldn't work normally.No attempts have been made to replace other accessories.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IPC® CONSOLE - LEGEND EHS
Common Device Name
DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE
Manufacturer (Section D)
MEDTRONIC XOMED INC.
6743 southpoint dr n
jacksonville FL 32216
Manufacturer (Section G)
MEDTRONIC XOMED INC.
6743 southpoint dr n
jacksonville FL 32216
Manufacturer Contact
glen belmer
6743 southpoint drive north
jacksonville, FL 32216
6122713209
MDR Report Key18409799
Report Number1045254-2023-01169
Device Sequence Number2690684
Product Code HBE
UDI-Device Identifier00643169588141
UDI-Public00643169588141
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K081475
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 12/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberEC300
Device Catalogue NumberEC300
Device Lot Number213315198
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/30/2023
Initial Report FDA Received Date12/28/2023
Date Device Manufactured05/12/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-