| Catalog Number |
2C8546 |
| Medical Device Problem Code |
Device Contamination with Chemical or Other Material (2944)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/05/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that a clearlink continu-flo solution set had dark particulate matter inside the tubing; further described as "something in the line." this occurred prior to patient use.There was no patient involvement.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
G1: device manufacturer address 1: 600 mts.Oeste de entrada, principal ave.Las americas, parque industrial should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: the actual device was not available; however, a photograph of the sample was provided for evaluation.A visual inspection of the photo with the unaided eye revealed a brown embedded particle in the tubing which resembled burned resin from the extrusion process.Due to the nature of the sample no further testing could be performed.The reported condition was verified.The cause of the condition was determined to be a manufacturing related issue; probably due to burned resin which can be generated in the cannon and accumulate on the pin and bushing combinations.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Search Alerts/Recalls
|