| Model Number |
WA47706S |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/11/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported to olympus, the resection electrode's loop became detached from the body of the electrode.The issue was found as the surgeon was about to initiate use on the patient for a hysteroscopic resection of fibroid.The device was removed from the surgical field before use on the patient.There were no reports of patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
To date, the device has not been returned to olympus for evaluation.The investigation is ongoing and follow up with the user facility is currently being performed.A supplemental report will be submitted upon completion of the investigation or if additional information is provided by the user facility.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the device evaluation and the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 1 year since the subject device was manufactured.The device was returned to olympus for inspection, and the customer's complaint was confirmed.Based on the results of the investigation, a definitive root cause could not be determined.However, it is likely that "loop electrode broke as the surgeon was about to initiate use on patient.One end of the loop became detached from the body of the electrode.Detected before use on patient and removed from surgical field" occurred due to improper handling.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
The customer reports the procedure was therapeutic.
|
| |
|