Model Number WA47706S
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/11/2023
Type of Reportable Event Malfunction
Event or Problem Description
The customer reported to olympus, the resection electrode's loop became detached from the body of the electrode.The issue was found as the surgeon was about to initiate use on the patient for a hysteroscopic resection of fibroid.The device was removed from the surgical field before use on the patient.There were no reports of patient harm.
 
Additional Manufacturer Narrative
To date, the device has not been returned to olympus for evaluation.The investigation is ongoing and follow up with the user facility is currently being performed.A supplemental report will be submitted upon completion of the investigation or if additional information is provided by the user facility.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the device evaluation and the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 1 year since the subject device was manufactured.The device was returned to olympus for inspection, and the customer's complaint was confirmed.Based on the results of the investigation, a definitive root cause could not be determined.However, it is likely that "loop electrode broke as the surgeon was about to initiate use on patient.One end of the loop became detached from the body of the electrode.Detected before use on patient and removed from surgical field" occurred due to improper handling.Olympus will continue to monitor field performance for this device.
 
Event or Problem Description
The customer reports the procedure was therapeutic.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HF-RESECTION ELECTRODE "PLASMALOOP", LOOP, MEDIUM, 24 FR., 12°-30°, ESG TCRIS
Common Device Name
RESECTION ELECTRODE
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18410173
Report Number9610773-2023-03792
Device Sequence Number8723510
Product Code FAS
UDI-Device Identifier14042761085530
UDI-Public14042761085530
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K171965
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 08/02/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberWA47706S
Device Lot Number1000118996
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/14/2023
Supplement Date Received by Manufacturer07/30/2024
Initial Report FDA Received Date12/28/2023
Supplement Report FDA Received Date08/02/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/17/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-