• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ENCORE 26; SYRINGE, BALLOON INFLATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ENCORE 26; SYRINGE, BALLOON INFLATION Back to Search Results
Model Number 25335
Medical Device Problem Code Device Contamination with Chemical or Other Material (2944)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/01/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
B3 date of event: date of event was approximated to (b)(6) /2023 as no event date was reported.
 
Event or Problem Description
It was reported that a foreign material was present in the device.An encore 26 inflation device was selected for use.During preparation, it was noted that the device contained what appears to be a hair.The procedure was completed with another of the same device.No patient injury reported.
 
Additional Manufacturer Narrative
B3 date of event: date of event was approximated to 12/01/2023 as no event date was reported.The device was returned for analysis.Visual inspection revealed no sign of hair or any silicon droplet.
 
Event or Problem Description
It was reported that a foreign material was present in the device.An encore 26 inflation device was selected for use.During preparation, it was noted that the device contained what appears to be a hair.The procedure was completed with another of the same device.No patient injury reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENCORE 26
Common Device Name
SYRINGE, BALLOON INFLATION
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
model farm road
cork T12 Y K88
EI   T12 YK88
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key18410433
Report Number2124215-2023-74180
Device Sequence Number8723515
Product Code MAV
UDI-Device Identifier08714729183624
UDI-Public08714729183624
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K955869
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/31/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/29/2025
Device Model Number25335
Device Catalogue Number25335
Device Lot Number0031930241
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/14/2023
Supplement Date Received by Manufacturer01/19/2024
Initial Report FDA Received Date12/28/2023
Supplement Report FDA Received Date01/31/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/30/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-