Catalog Number 200009902
Medical Device Problem Code Device Difficult to Setup or Prepare (1487)
Health Effect - Clinical Codes Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/04/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Analysis results are not available at the time of this report.A follow-up report will be sent when the analysis is complete.
 
Event or Problem Description
During an inbone total ankle replacement procedure, the 18mm base stem failed to thread into the other implants due to imperfections in its threads.Despite this setback, the surgery was successful as a backup 18mm base was used instead.The returned product will undergo examination to identify the flaw in the base stem.
 
Additional Manufacturer Narrative
The reported event was confirmed since the device was returned for evaluation and matches the alleged failure mode.The device inspection revealed the following: visual examination of the returned device confirms the size and lot number.Scratches can be seen on the inferior surface of the device.The external threads of the device were confirmed to be defective upon failing a feature check with a thread gage.The device could not pass dimensional verification as it was unable to thread into the thread gage.Based on investigation, the root cause was attributed to a manufacturing related issue.A nonconformance was opened to further address the issue.
 
Event or Problem Description
During an inbone total ankle replacement procedure, the 18mm base stem failed to thread into the other implants due to imperfections in its threads.Despite this setback, the surgery was successful as a backup 18mm base was used instead.The returned product will undergo examination to identify the flaw in the base stem.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INBONE STEM TIBIAL BASE SIZE 18MM RIGHT AND LEFT
Common Device Name
PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
WRIGHT MEDICAL TECHNOLOGY INC
1023 cherry rd
memphis TN 38117
Manufacturer (Section G)
WRIGHT MEDICAL TECHNOLOGY, INC.
11576 memphis arlington rd
arlington TN 38002
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key18411248
Report Number3010667733-2023-00815
Device Sequence Number15630964
Product Code HSN
UDI-Device Identifier00840420123848
UDI-Public00840420123848
Combination Product (Y/N)N
Initial Reporter StateSD
Initial Reporter CountryUS
PMA/510(K) Number
K051023
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number200009902
Device Lot Number1751953
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/04/2023
Supplement Date Received by Manufacturer02/23/2024
Initial Report FDA Received Date12/28/2023
Supplement Report FDA Received Date03/19/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/29/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-