C.R. BARD, INC. (BASD) -3006260740 POWERLOC SAFETY INFUSION SET 19G X 1 IN WITH Y-SITE; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Model Number N/A
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/07/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported by the customer "i have a port needle with y-site that leaked at the connector y-site.This area has notable white powder on the tubing and connector and is not site where the chartwell pump tubing was connected to." no other information was provided.
 
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The manufacturer has received the sample and is pending evaluation.Results are expected soon.
 
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The complaint of a leaking y-site was confirmed.The product returned for evaluation was one 19g powerloc infusion set w/y-site.A gross visual inspection of the returned unit found that two injection caps were returned attached to the two luers of the infusion set.The unit was leak tested by flushing water through each of the injection caps using a 12ml luer lok syringe.During testing, no leaks were observed.The same leak test was performed, except with the clamps engaged to simulate pressurization.A leak was then observed between the injection cap and the luer of the y-site.The injection cap was inspected and found to be loosely attached to the luer.The injection cap was tightened and the pressurized leak test performed again.No leaks were observed.Finally the unit was leak tested with and without pressurization by connecting the luer lok syringe directly to the luer of the infusion set and no leaks were observed again.Based on the available evidence, the reported leak was present but was caused by inadequate tightening of the injection cap.
 
Event or Problem Description
It was reported by the customer "i have a port needle with y-site that leaked at the connector y-site.This area has notable white powder on the tubing and connector and is not site where the chartwell pump tubing was connected to." no other information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERLOC SAFETY INFUSION SET 19G X 1 IN WITH Y-SITE
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer Contact
kayla bevins
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key18412788
Report Number3006260740-2023-05908
Device Sequence Number2673539
Product Code FPA
UDI-Device Identifier00801741047893
UDI-Public(01)00801741047893
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K060812
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/18/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number0671910
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/20/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/07/2023
Supplement Date Received by Manufacturer03/18/2024
Initial Report FDA Received Date12/28/2023
Supplement Report FDA Received Date03/27/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-