| Model Number |
DEPTH DEV,2.7/3.5 /4.0MMLO-PRO |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/30/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2023, it was reported by a arthrex employee via (b)(4).That an ar-8943-15 depth device stick portion is broken.This was discovered during use in a case with no patient effect.
|
| |
|
Additional Manufacturer Narrative
|
|
Complaint allegation is confirmed.Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause of this failure is attributed to wear and tear damage incurred over repeated usage.
|
| |
|