• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN LP BARRX BARRX; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN LP BARRX BARRX; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number 64082
Medical Device Problem Codes Inflation Problem (1310); Material Twisted/Bent (2981)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/30/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during the rfa (radio frequency ablation) procedure for barrett¿s esophagus, the balloon was filled with air, but would not achieve the proper size which signals the physician to do the ablation.The catheter was removed and a defect was noticed.Ablation was never delivered with the defective device.The coil was bent inward which would block the balloon from achieving full volume.A different catheter was used to perform the case.There was no patient or user harm.
 
Additional Manufacturer Narrative
Evaluation summary: medtronic conducted an investigation based upon all information received.The device and a photo were available for evaluation.Visual inspection noted no abnormalities.The connector pins were intact.The electrode was not torn.The wire that attaches the electrode is twisted as if the electrode was installed backwards.Functionally, the balloon was manually inflated with a syringe and able to hold pressure.The catheter electrically erasable programmable read-only memory (eeprom) was read and e96 error was observed.The catheter was connected to a halo flex generator and recognized as new.When inflation was attempted, the balloon never fully inflated to the point it was ready for ablation.An air leak error was displayed while attempting to inflate the catheter.The balloon was submerged in and manually inflated and held air, no air bubbles were noted.It was reported that the catheter coils are kinked and the catheter balloon failed to inflate.The reported issues were confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARRX
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
COVIDIEN LP BARRX
covidien gi solutions
sunnyvale CA 94085 4022
Manufacturer (Section G)
COVIDIEN LP BARRX
covidien gi solutions
sunnyvale CA 94085 4022
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key18414003
Report Number3004904811-2023-00020
Device Sequence Number21010203
Product Code GEI
UDI-Device Identifier10884521211599
UDI-Public10884521211599
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K142364
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/06/2023
Device Model Number64082
Device Catalogue Number64082
Device Lot NumberB000002333
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/18/2023
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/30/2023
Supplement Date Received by Manufacturer05/16/2024
Initial Report FDA Received Date12/28/2023
Supplement Report FDA Received Date05/16/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/06/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-