| Model Number |
64082 |
| Medical Device Problem Codes |
Inflation Problem (1310); Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/30/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during the rfa (radio frequency ablation) procedure for barrett¿s esophagus, the balloon was filled with air, but would not achieve the proper size which signals the physician to do the ablation.The catheter was removed and a defect was noticed.Ablation was never delivered with the defective device.The coil was bent inward which would block the balloon from achieving full volume.A different catheter was used to perform the case.There was no patient or user harm.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: medtronic conducted an investigation based upon all information received.The device and a photo were available for evaluation.Visual inspection noted no abnormalities.The connector pins were intact.The electrode was not torn.The wire that attaches the electrode is twisted as if the electrode was installed backwards.Functionally, the balloon was manually inflated with a syringe and able to hold pressure.The catheter electrically erasable programmable read-only memory (eeprom) was read and e96 error was observed.The catheter was connected to a halo flex generator and recognized as new.When inflation was attempted, the balloon never fully inflated to the point it was ready for ablation.An air leak error was displayed while attempting to inflate the catheter.The balloon was submerged in and manually inflated and held air, no air bubbles were noted.It was reported that the catheter coils are kinked and the catheter balloon failed to inflate.The reported issues were confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|