WRIGHT MEDICAL TECHNOLOGY INC INFINITY ADAPTIS TALDOME SZ1 INFINITY ADAPTIS; PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Catalog Number 33680021
Medical Device Problem Codes Loss of Osseointegration (2408); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Inadequate Osseointegration (2646); Insufficient Information (4580)
Date of Event 12/04/2023
Type of Reportable Event Serious Injury
Event or Problem Description
The digital stryker prophecy team received a ct scan indicating that the patient require a revision surgery of the tibial and talar components.The surgery won't be done until at least february.
 
Additional Manufacturer Narrative
Based on the available information the device will not be returned therefore an evaluation of the device cannot be performed.Should additional information become available, it will be provided in a supplemental report.H3 other text : device disposition unknown.
 
Event or Problem Description
The digital stryker prophecy team received a ct scan indicating that the patient require a revision surgery of the tibial and talar components.The surgery won't be done until at least february.
 
Additional Manufacturer Narrative
Please note correction to h6 (device and health impact codes).The reported event could be confirmed since images of ct scans were provided and shows loosening of the talar component.Upon further investigation of the ct scans by healthcare professionals the following was observed: ¿the talus shows some radiolucence as well, anteriorly and a little bit around the pegs.Loosening is likely, there is no clear sign of migration.¿ based on investigation, the root cause was attributed to a patient related issue.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If the device is returned or if any additional information is provided, the investigation will be reassessed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFINITY ADAPTIS TALDOME SZ1 INFINITY ADAPTIS
Common Device Name
PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
WRIGHT MEDICAL TECHNOLOGY INC
1023 cherry rd
memphis TN 38117
Manufacturer (Section G)
WRIGHT MEDICAL TECHNOLOGY INC
1023 cherry rd
memphis TN 38117
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key18414207
Report Number3010667733-2023-00816
Device Sequence Number2748446
Product Code HSN
UDI-Device Identifier00889797101516
UDI-Public00889797101516
Combination Product (Y/N)N
Initial Reporter StateNH
Initial Reporter CountryUS
PMA/510(K) Number
K172633
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/25/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Catalogue Number33680021
Device Lot Number1697734
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/04/2023
Supplement Date Received by Manufacturer02/29/2024
Initial Report FDA Received Date12/28/2023
Supplement Report FDA Received Date03/25/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/14/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-