| Model Number |
6131 |
| Medical Device Problem Codes |
Failure to Deliver Energy (1211); Failure to Cut (2587)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/07/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that a captivator snare was used in the colon to resect a polyp during a colonoscopy procedure performed on (b)(6) 2023.During the procedure, the snare was plugged into the generator for energization.The snare was then looped around the polyp and physician stepped on the blue pedal to activate the energy and cauterize the polyp.However, the snare did not deliver energy and no cauterization was made.No visible problem was noted with the cautery pin and handle.The procedure was completed with another captivator snare.There were no patient complications reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Block h6: imdrf device code a090402 captures the reportable event of snare unable to deliver energy.Imdrf device code a050702 captures the reportable event of snare loop cutting problems.
|
| |
|