| Model Number |
72003 |
| Medical Device Problem Code |
Failure to Advance (2524)
|
| Health Effect - Clinical Code |
Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
11/28/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Patient has a history of other failed surgical interventions which the patient reports are due to scar tissue.Medical records for those failed attempts are not available to the nalu firm for review and confirmation.Patient is reported to have a congenitally narrow canal and likely significant spinal stenosis, making entry and placement of the nalu leads difficult and leading to the procedure being aborted.There are no reports from the implanting physician of any surgical tool failure or nalu system or component failure.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2023 a surgical procedure was attempted and aborted.Patient was scheduled to receive the nalu spinal cord stimulator trial system implanted, however upon attempting to insert the leads the physician found he was unable to insert the leads past the t10 vertebra.Physician attempted entry at l1/2 and moved up to t12/l1 also with no success.Physician injected contrast dye into the epidural space and was not able to achieve any flow upward or laterally.Ultimately the physician elected to abort the procedure without placing the system.
|
| |
|