| Model Number |
SONATA |
| Medical Device Problem Codes |
Circuit Failure (1089); Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/08/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
When the user was followed up, it was not possible to visualize esrt responses unlike in previous exams in which responses were recorded on all channels.The neural responses (art) also seemed to have an atypical morphology, and during tests, maestro signalled several times about the compliance/load on the electrodes.These observed changes and a channel with abnormal impedance caught the attention of the implant team.As per additional information the recipient keeps atypical neural responses and variations between the outcomes.Reportedly, there is no evolution in oral language development as well as in auditory skills as the recipient only detects sounds.Additional measurements performed on 16-oct-2023 have been reviewed by clinical support and it was indicated that the device is no longer working within specifications.Reimplantation is now considered.
|
| |
|
Additional Manufacturer Narrative
|
|
The device has not been explanted.If it should be explanted, it should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
|
| |
|
Event or Problem Description
|
|
When the user was followed up, it was not possible to visualize esrt responses unlike in previous exams in which responses were recorded on all channels.The neural responses (art) also seemed to have an atypical morphology, and during tests, maestro signaled several times about the compliance/load on the electrodes.These observed changes and a channel with abnormal impedance caught the attention of the implant team.As per additional information, the recipient has atypical neural responses and variations between the outcomes.Reportedly, there is no evolution in oral language development as well as in auditory skills as the recipient only detects sounds.Additional measurements performed on (b)(6) 2023 have been reviewed by clinical support and it was indicated that the device is no longer working within specifications.Re-implantation is being considered.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: based on the received in situ measurements from the field, a technical implant failure is likely.In addition, a partial migration of the active electrode is confirmed, which might have contributed to the reported issues in performance.However, to determine an exact root cause an investigation of the explanted device would be necessary.Re-implantation is considered, but no date has been scheduled yet.
|
| |
|
Additional Manufacturer Narrative
|
|
Based on the received in situ measurements from the field, a technical implant failure is likely.In addition, a partial migration of the active electrode is confirmed, which might have contributed to the reported issues in performance.However, to determine an exact root cause an investigation of the explanted device would be necessary.Re-implantation was carried out, but the device has not been received for investigation yet.
|
| |
|
Event or Problem Description
|
|
When the user was followed up, it was not possible to visualize esrt responses unlike in previous exams in which responses were recorded on all channels.The neural responses (art) also seemed to have an atypical morphology, and during tests, maestro signaled several times about the compliance/load on the electrodes.These observed changes and a channel with abnormal impedance caught the attention of the implant team.As per additional information, the recipient has atypical neural responses and variations between the outcomes.Reportedly, there is no evolution in oral language development as well as in auditory skills as the recipient only detects sounds.The recipient has been re-implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Conclusion: device investigation confirmed that stimulator electronics is working according to specifications.However, a reduced electrical impedance between coil and eap/rg (evoked action potentials/ remote ground) electrode was discovered during investigation, which very likely caused the reported issues.In addition, a partial migration of the active electrode is confirmed, which might have contributed to the reported issues in performance.Other mechanical damages found during investigation are attributable to the explantation surgery.This is a final report.
|
| |
|
Event or Problem Description
|
|
During a follow up, it was not possible to visualize esrt responses unlike in previous exams in which responses were recorded on all channels.The neural responses (art) also seemed to have an atypical morphology, and during tests, maestro signaled several times about the compliance/load on the electrodes.These observed changes and a channel with abnormal impedance caught the attention of the implant team.As per additional information, the recipient has atypical neural responses and variations between the outcomes.Reportedly, there is no evolution in oral language development as well as in auditory skills as the recipient only detects sounds.The recipient has been re-implanted.
|
| |
|