Model Number N/A
Medical Device Problem Code Difficult to Insert (1316)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/14/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a tasp was returned fractured.Attempts to obtain additional information have been made; however, no more information is available.
 
Additional Manufacturer Narrative
(b)(4).Visual examination of the returned product shows that the product exhibits signs of repeated use (nicks and gouges) and the dovetail is fractured.All pieces were not returned.Medical records were not provided.Device history record (dhr) was reviewed for deviations and/ or anomalies with no deviations / anomalies identified.A definitive root cause cannot be determined.However, the ztr wa 0820 20 documents the summary of investigations for complaints associated with the tibial articular surface provisional (tasp) fracture.Fractured tasps were analyzed by optical microscopy revealing hackle marks and river lines in the case of bending overload; and hackle marks, river lines and striations were found in the case of low cycle fatigue culminating in a bending overload failure type.Complaint is confirmed via product return and evaluation.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.H6: component code: proposed component (annex g) code is mechanical (g04) - provisional top.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PS TIBIAL ARTICULAR SURFACE PROVISIONAL RIGHT SIZE 6-9 CD TOP
Common Device Name
INSTRUMENT, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key18418147
Report Number0001822565-2023-03796
Device Sequence Number20824444
Product Code JWH
UDI-Device Identifier00889024245310
UDI-Public(01)00889024245310(10)63350614
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K113369
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 12/29/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number42527400510
Device Lot Number63350614
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/27/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/22/2023
Initial Report FDA Received Date12/29/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/04/2016
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Patient SexUnknown
-
-