Model Number N/A
Medical Device Problem Codes Device Dislodged or Dislocated (2923); Naturally Worn (2988)
Health Effect - Clinical Codes Joint Dislocation (2374); Osteolysis (2377); Osteopenia/ Osteoporosis (2651); Metal Related Pathology (4530)
Date of Event 12/07/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Multiple mdr reports were filed for this event, please see associated reports: 0001822565 - 2023 - 03824, 0001822565 - 2023 - 03826.The customer has indicated that the product will not be returned to zimmer biomet for investigation as it was not returned by the hospital.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that the patient was revised approximately twelve years post-implantation due to a dislocation.During the revision surgery, the surgeon noticed trunnion wear and a pseudotumor possibly due to metal ion release.It was reported that no further information is available.
 
Additional Manufacturer Narrative
(b)(4).This follow-up report is being submitted to relay additional information.The following sections were updated/corrected: updated: g3; h2; h3; h6 h6: proposed component code: mechanical (g04)- head.Visual examination of the provided pictures identified black debris on the inside of the head as well as on the trunnion of the stem.Both devices are covered in bio-debris, and the picture of the stem shows it still remains within the joint.The surrounding tissues are discolored.No other damage was noted.Review of the device history records identified no deviations or anomalies during manufacturing.Radiographs were provided and reviewed.Review of the available records identified the following: right hip arthroplasty dislocation.Osteolysis as noted.A definitive root cause cannot be determined.This complaint was confirmed based on the provided x-rays confirming the dislocation and provided pictures of the implants.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further information is available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FEMORAL HEAD STERILE PRODUCT DO NOT RESTERILIZE 12/14 TAPER
Common Device Name
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER MANUFACTURING B.V.
turpeaux industrial park
route #1 km 123.4 bldg #1
mercedita PR 00715
Manufacturer (Section G)
ZIMMER MANUFACTURING B.V.
turpeaux industrial park
route #1 km 123.4 bldg #1
mercedita PR 00715
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key18418451
Report Number0002648920-2023-00324
Device Sequence Number18109613
Product Code LPH
UDI-Device Identifier00889024144668
UDI-Public(01)00889024144668(17)210228(10)61725732
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K953337
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2011
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/20/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2021
Device Model NumberN/A
Device Catalogue Number00801803201
Device Lot Number61725732
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/07/2023
Supplement Date Received by Manufacturer05/14/2024
Initial Report FDA Received Date12/29/2023
Supplement Report FDA Received Date05/20/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/16/2011
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexFemale
Patient Weight84 KG
-
-