MAUDE Adverse Event Report: STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC INTRAOCULAR LENS Back to Search Results
Model Number VICM5_13.2
Medical Device Problem Code Inadequacy of Device Shape and/or Size (1583)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/15/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Claim# (b)(4).
 
Event or Problem Description
The reporter indicated that a 13.2mm vicm5_13.2 implantable collamer lens of -3.50 diopter was implanted into the patients right eye (od) on (b)(6) 2023.On (b)(6) 2023 lens repositioning was performed due to excessie vault but the problem did not resolve.On (b)(6) 2023 the lens was exchanged for a shorter length lens.This exchange resolved the problem.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE COLLAMER LENS (ICL)
Common Device Name
PHAKIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer (Section G)
STAAR SUGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer Contact
joselene muniz
1911 walker avenue
monrovia, CA 91016
MDR Report Key18418705
Report Number2023826-2023-05829
Device Sequence Number12729497
Product Code MTA
UDI-Device Identifier00840311300808
UDI-Public00840311300808
Combination Product (Y/N)N
Initial Reporter CountrySZ
PMA/510(K) Number
P030016
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 12/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/31/2026
Device Model NumberVICM5_13.2
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/15/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/05/2023
Initial Report FDA Received Date12/29/2023
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/25/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CARTRIDGE MODEL: SFC-45 W/FTP, LOT# UNK; INJECTOR MODEL: MSI-PF, LOT# UNK
Outcome Attributed to Adverse Event Required Intervention;
Patient Age35 YR
Patient SexFemale
-
-