• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERRAGENE S.A. STRYKER SLCD12; AUTO-READER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERRAGENE S.A. STRYKER SLCD12; AUTO-READER Back to Search Results
Model Number STRYKER SLCD12
Medical Device Problem Code Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/24/2022
Type of Reportable Event Malfunction
Event or Problem Description
According to the distributor representative, the autoreader incubator would not boot up, since it was stuck in the loading screen.
 
Additional Manufacturer Narrative
No device-related deaths or serious injuries were informed.Device malfunction cause the device to fail to perform its essential function and compromise the device's monitoring effectiveness, since it would not be able to process biological indicators.If this were to recur, the lack of biological indicators processing could affect the load release, which could contribute to a serious injury due to lack of sterile material.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STRYKER SLCD12
Common Device Name
AUTO-READER
Manufacturer (Section D)
TERRAGENE S.A.
ruta nac. no 9, km 280
parque industrial micropi
alvear, santa fe 2130
AR  2130
Manufacturer (Section G)
TERRAGENE S.A.
ruta nac. no 9, km 280
parque industrial micropi
alvear, santa fe 2130
*   2130
Manufacturer Contact
hernando carrizo
ruta nac. n[?] 9, km 280
parque industrial micropi
alvear, santa fe 2130
AR   2130
MDR Report Key18418819
Report Number3013145340-2023-00016
Device Sequence Number20553753
Product Code FRC
UDI-Device Identifier07798164678175
UDI-Public07798164678175
Combination Product (Y/N)N
PMA/510(K) Number
K191021
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 12/29/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/22/2025
Device Model NumberSTRYKER SLCD12
Device Lot NumberSLCD124820
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/24/2022
Initial Report FDA Received Date12/29/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/22/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-