| Catalog Number |
72202602 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/08/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H10: internal complaint reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that during an arthroscopy, the twinfix ultra anchor broke.The procedure was completed using a smith and nephew backup device; however, it is unknown if there was a delay.No further complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: internal complaint reference: (b)(4).H3, h6: the reported device was received for evaluation.A visual inspection of the returned device found that it is not in its original packaging.The insertion device was returned with no suture material or anchor fragments.The distal tip of the insertion device has been fractured away and was not returned.Based on the condition of the product material found during visual inspection, additional material testing is not required.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A review of the anchor material specifications found that the storage requirements for the material are specified, and the material must comply with provided percentage composition specifications as measured by ash test.The root cause has been associated with unintended use of the device.Factors that could have contributed to the failure include excessive force on the device, excessive torque on the device, attempted correction of a damaged device, off-axis insertion, improper preparation of the insertion site, or an inadvertent impact event inconsistent with normal use.No containment or corrective actions are recommended at this time.
|
| |
|
Event or Problem Description
|
|
It was reported that during an arthroscopy, the twinfix ultra anchor broke.It is unknown if the event happened internal or external to patient.The procedure was completed using a smith and nephew backup device; however, it is unknown if there was a delay.No further complications were reported.
|
| |
|
Search Alerts/Recalls
|