| Catalog Number |
1651-38 |
| Medical Device Problem Code |
Failure to Deflate (4060)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/01/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The catheter was prepared and inserted into the vessel.The balloon was then inflated with the diluted contrast medium.During the thrombectomy, an attempt was made to deflate the balloon, but it was difficult to deflate as it took longer than usual.Therefore, this catheter was no longer used and a new catheter of the same size was prepared.The operation was successfully completed with replacement catheter.No injury was reported to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
We have received the device for investigation.The reported problem was confirmed.During testing the balloon failed to deflate.When filling/attempting to deflate the balloon resistance was felt indicating a blockage in the inflation path.When the lumen of the device was cut approximately half way along the catheter the blockage was found and fluid was able to pass through the remaining portion of the lumen.The exact source of the blockage in the inflation lumen could not be identified.We have conducted a lot history review and did not find any issues noted in the manufacturing or packaging process that could be related to this issue.All qc tests passed their requirements.Further, we have not received any other complaints of a similar nature for devices from this lot.
|
| |
|