| Model Number |
VTICM5_13.7 |
| Medical Device Problem Code |
No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Code |
Corneal Edema (1791)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The reporter indicated that a 13.7mm vticm5_13.7 implantable collamer lens of -8.50/1.0/147 (sphere/cylinder/axis) was implanted into the patients right eye (od) on (b)(6) 2023 as a replacement lens to address low vault with rotation.For status of the eye the reporter stated that the eye was first day postoperative with corneal edema and miosis from surgery.It was reported that at the next appointment on (b)(6) 2023 pupil was still in miosis from myostst.Additional information has been requested but none has been forthcoming.If additional information is received a supplemental medwatch report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
H6- health effect- clinical code 4581: miosis.H6 type of investigation lens work order search: no similar complaint type events reported for units within the same lot.Claim# (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
The reporter indicated the surgeon implanted a 13.7mm vticm5_13.7 implantable collamer lens of diopter -8.5/+1.0/130 (sphere/cylinder/axis) into the patient's right eye (od) (b)(6) 2023 as a replacement lens.The patient experienced corneal oedema, miosis and probable low vault.Reportedly "eye was first day postoperative with corneal edema and miosis from surgery" and "pupil still in miosis from myostat, which is also the probable explanation of the low vault".The lens remains implanted.Claim# (b)(4).
|
| |
|