| Model Number |
MAJ-824 |
| Medical Device Problem Code |
Improper or Incorrect Procedure or Method (2017)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/29/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device has not been returned for evaluation.Photographs of the service inspection were provided.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided.
|
| |
|
Event or Problem Description
|
|
The customer reported that the reprocessor gave an error indicating an irregularity in circuitry (e83).During inspection of the reprocessor, examination showed the water filter (part number maj-824) had not been replaced since 2021.The device instructions for use for the reprocessor states monthly replacement of this filter is required.No adverse effects to patient reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.The device history record was unable to be reviewed for this device since the serial number was not provided.However, olympus only releases products to market that meet all manufacturing specifications and final product release criteria.Based on the results of the investigation, since the user did not read the instructions for use carefully, they incorrectly managed water filter replacement and used the filter even after its replacement period has expired.The event can be prevented by following the instructions for use (ifu) which state: "oer-aw operation manual.Chapter 7 routine maintenance.7.2 replacing the water filter (maj-824).Replace the water filter at least once a month to prevent contamination of the rinse water.The water filter should also be replaced whenever an error code indicating water supply insufficiency [e01] is displayed.The water filter should be replaced by following the flow shown below.".Olympus will continue to monitor field performance for this device.
|
| |
|
Search Alerts/Recalls
|