Model Number MAJ-824
Medical Device Problem Code Improper or Incorrect Procedure or Method (2017)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/29/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device has not been returned for evaluation.Photographs of the service inspection were provided.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided.
 
Event or Problem Description
The customer reported that the reprocessor gave an error indicating an irregularity in circuitry (e83).During inspection of the reprocessor, examination showed the water filter (part number maj-824) had not been replaced since 2021.The device instructions for use for the reprocessor states monthly replacement of this filter is required.No adverse effects to patient reported.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.The device history record was unable to be reviewed for this device since the serial number was not provided.However, olympus only releases products to market that meet all manufacturing specifications and final product release criteria.Based on the results of the investigation, since the user did not read the instructions for use carefully, they incorrectly managed water filter replacement and used the filter even after its replacement period has expired.The event can be prevented by following the instructions for use (ifu) which state: "oer-aw operation manual.Chapter 7 routine maintenance.7.2 replacing the water filter (maj-824).Replace the water filter at least once a month to prevent contamination of the rinse water.The water filter should also be replaced whenever an error code indicating water supply insufficiency [e01] is displayed.The water filter should be replaced by following the flow shown below.".Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WATER FILTER
Common Device Name
WATER FILTER
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18419655
Report Number9610595-2023-20583
Device Sequence Number18133795
Product Code FEB
UDI-Device Identifier04953170069482
UDI-Public04953170069482
Combination Product (Y/N)N
Initial Reporter CountryEC
PMA/510(K) Number
K103264
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 02/06/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMAJ-824
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/30/2023
Supplement Date Received by Manufacturer01/31/2024
Initial Report FDA Received Date12/29/2023
Supplement Report FDA Received Date02/06/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
OER-AW ENDOSCOPE REPROCESSOR
Patient SexUnknown
-
-