| Catalog Number |
100/905/340-CH |
| Medical Device Problem Code |
Labelling, Instructions for Use or Training Problem (1318)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
09/18/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that "single limb ventilator circuit connected to a ventilator requiring use of a two limbed ventilator tubing circuit.Packaging lacked warning and ventilator able to deliver inspiration but did not allow for exhalation.The ventilator model used was the zoll eagle ii which requires the 2-limb circuitry tubing; however, the therapist utilized the single limb circuit tubing intended for use with the parapac transport ventilator.There was temporary harm ¿ the insertion of a chest tube".
|
| |
|
Additional Manufacturer Narrative
|
|
D4: expiration date and h4: manufacture date is unknown; no information is available based on reported lot number.H3: other; device not returned to manufacturer.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
| |
|
Additional Manufacturer Narrative
|
|
H6.Investigation codes: updated.Investigation summary: no product sample was received; therefore, visual and functional testing could not be performed.The reported issue could not be confirmed.If the product is returned, the manufacturer will reopen this complaint for further investigation.A device history record could not be completed as no lot number was received.
|
| |
|
Additional Manufacturer Narrative
|
|
On review of the file, it was found that it was entered on the incorrect record type.On further review it was determined that the device does not have a 510k or sales in the united states.Nor are there any substantially similar devices sold in the us.Therefore, no regulatory reporting is required.Please disregard any reports associated with file mrn 3012307300-2023-12280.
|
| |
|