| Catalog Number |
109990 |
| Medical Device Problem Code |
Disconnection (1171)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that the access (red stripped) line of five (5) prismaflex m150 sets disconnected from below the red coupling which resulted in an external blood leak.These connection issues occurred during treatment with continuous renal replacement therapy (crrt).There was no report of patient injury or medical intervention associated with this event.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: the actual device was not available; however, a photograph of the sample was provided for evaluation.During visual inspection of the photo, it was observed that the luer connector was disconnected from the tubing which allowed leakage.The lines of the set assembled with the luer lock are provided by a baxter supplier.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.The reported condition was verified.The cause of the tubing disconnection was determined to be related to a gluing issue in the manufacturing process.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|