• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TREACE MEDICAL CONCEPTS AO HEXALOBULAR #8 TIP; DRIVER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TREACE MEDICAL CONCEPTS AO HEXALOBULAR #8 TIP; DRIVER Back to Search Results
Model Number 1405-2202
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 12/01/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the driving end of an ao driver tip broke off as the surgeon was inserting a screw.The detached portion of the ao driver tip was retained in the head of a screw, which was implanted in the patient.Although there was no adverse event reported or anticipated, a piece of an ao tip is not intended to be implanted, therefore the company has chosen to file a mdr to ensure full compliance with 21 cfr part 803.
 
Additional Manufacturer Narrative
It was reported that the driving end of an ao driver tip broke off as the surgeon was inserting a screw.The detached portion of the ao driver tip was retained in the head of a screw, which was implanted in the patient.A backup device was available to successfully complete the surgery.No other patient outcomes were reported.No device was returned for evaluation.Device specific information was not available; therefore, a review of device history records was not able to be performed.However, all non-conformances for possible driver tips utilized in surgery were reviewed and no non-conformances or issues during the manufacture or release of the products were identified to date that could have contributed to what was reported.The most likely cause cannot be determined due to the device not being returned.However, it is likely that excessive torque was applied to the device during use, which can lead to torsional failure.The surgical technique includes multiple statements regarding the proper method for inserting screws into a plate.Although the company has determined that the subject event in this mdr is unlikely to result in a serious injury if it recurred, the company has decided to file this mdr in an abundance of caution and to ensure full compliance with 21 cfr part 803.The company will supplement this mdr as necessary and appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AO HEXALOBULAR #8 TIP
Common Device Name
DRIVER
Manufacturer (Section D)
TREACE MEDICAL CONCEPTS
100 palmetto park place
ponte vedra FL 32081
Manufacturer (Section G)
TREACE MEDICAL CONCEPTS
100 palmetto park place
ponte vedra FL 32081
Manufacturer Contact
kristine gates
100 palmetto park place
ponte vedra, FL 
9043735940
MDR Report Key18421973
Report Number3011623994-2023-00348
Device Sequence Number20824476
Product Code HXX
UDI-Device Identifier00853114006624
UDI-Public00853114006624
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
N/A-CLASS I
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 12/29/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1405-2202
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/01/2023
Initial Report FDA Received Date12/29/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-