• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NUVASIVE INC NUVASIVE ACP SYSTEM; APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NUVASIVE INC NUVASIVE ACP SYSTEM; APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY Back to Search Results
Model Number 18008515
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Codes Laceration(s) of Esophagus (2398); Perforation of Esophagus (2399); Swelling/ Edema (4577)
Date of Event 12/08/2023
Type of Reportable Event Serious Injury
Event or Problem Description
On (b)(6) 2023 a patient underwent a 2-level anterior cervical "discectomy" fusion.About 3 months post-op the patient went to the er due to fluid build up.On (b)(6) 2023 a revision surgery was conducted where it was discovered a screw backed out of the plate causing a small esophageal tear.The backed out screw was replaced and a c4/7 posterior cervical fusion was conducted.No additional issues reported.
 
Additional Manufacturer Narrative
No product was returned, but images provided confirmed the complaint.The patient's post-op physical activity is unknown and it is unknown if the patient suffered a fall.The patient's bone quality is unknown.Review of the provided radiograph identified one of the inferior fixation screws backed out approximately 10mm and the two superior fixation screws show excessive angulation and a proud screw head.No definitive root cause can be determined however review of the information provided suggests plate and screw positioning, insufficient lock engagement, excessive screw angulation, and or excessive post-op physical activity as possible cause or contributors.Should more information be obtained a follow up report will be submitted.Labeling review: "potential adverse events and complications as with any major surgical procedures, there are risks involved in orthopedic surgery.Potential risks identified with the use of this system, which may require additional surgery, include: bending, fracture or loosening of implant component(s)." "warnings, cautions and precautions the implantation of spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.Proper patient selection is critical to the success of the procedure.Only patients who satisfy the criteria set forth under the indications section of this document and who do not have any of the conditions set forth under the contraindications section of this document should be considered for spinal fixation surgery using the nuvasive acp system.In addition, patients who smoke have been shown to have an increased incidence of pseudarthrosis.Based upon the fatigue testing results, the physician/surgeon should consider the levels of implantation, patient weight, patient activity level, other patient conditions, etc, which may impact the performance of the system.Follow proper instrument selection as specified in the surgical technique.The nuvasive acp system is designed to provide biomechanical stabilization as an adjunct to cervical fusion and should be used with anterior column support.Without anterior column support, its use may not be successful.Spinal fixation should only be undertaken after the surgeon has had hands on training in this method of spinal fixation and has become thoroughly knowledgeable about spinal anatomy and biomechanics.A surgical technique is available for instructions on the important aspects of this surgical procedure.Care should be taken to ensure that all components are ideally fixated prior to closure." "patient education: preoperative instructions to the patient are essential.The patient should be made aware of the limitations of the implant and potential risks of the surgery.The patient should be instructed to limit postoperative activity, as this will reduce the risk of bent, broken or loose implant components.The patient must be made aware that implant components may bend, break or loosen even though restrictions in activity are followed." "pre-operative warnings only patients that meet the criteria described in the indications should be selected.2.Patient condition and/or predispositions such as those addressed in the aforementioned contraindications should be avoided.3.Care should be used in the handling and storage of the acp implants.The implants should not be scratched or damaged.Implants and instruments should be protected during storage and from corrosive environments.For sterile implants: assure highly aseptic surgical conditions, and use aseptic technique when removing the acp implant from its packaging.Inspect the implant and packaging for signs of damage, including scratched or damaged devices or damage to the sterile barrier.Do not use the acp implants if there is any evidence of damage.Care should be used during surgical procedures to prevent damage to the device(s) and injury to the patient." "post-operative warnings damage to the weight-bearing structures can give rise to loosening of the components, dislocation and migration, as well as to other complications.To ensure the earliest possible detection of such catalysts of device dysfunction, the devices must be checked periodically postoperatively, using appropriate radiographic techniques." "method of use please refer to the surgical technique for this device" "information to obtain a surgical technique manual or should any information regarding the products or their uses be required, please contact your local representative or nuvasive directly at +1-800-475-9131.You may also email: info@nuvasive.Com." 9400042-en f-2023-05.
 
Event or Problem Description
It was reported that the patient underwent an initial two (2) level anterior cervical discectomy and fusion (acdf) procedure at c4-c6.Subsequently, approximately 3 months later, the patient was seen in the er for fluid build up.Imaging of the patient noted a screw had backed out of the plate causing a small esophageal tear.The patient was revised by removing the backed-out screw and adding c4-c7 posterior cervical fusion.The plate was retained as the ent approach surgeon believed removal would have been too risky.No further patient impact was reported.No additional information is available.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The reported event was confirmed by review of the radiograph provided, showing one of the screws in the c6 vertebral body, at the caudal end of the two (2) level plate, had backed out of the plate construct approximately 10mm.No product was returned to nuvasive for evaluation; further, operative notes and/or immediate post-op radiograph images were not provided for review of usage/technique.It is unknown if the construct was properly positioned and locked during the initial procedure.The patient's bone quality, post-op activity levels, and whether they experienced any falls or other trauma was not provided.A review of manufacturing records was unable to be performed as the lot information of the product involved in the event was not available.A definitive root cause was unable to be determined with the information provided; however, may be related to insufficient positioning of the screw, plate, and/or lock component or failure of the plate locking component.Labeling review: ".Potential adverse events and complications: as with any major surgical procedures, there are risks involved in orthopedic surgery.Potential risks identified with the use of this system, which may require additional surgery, include: bending, fracture or loosening of implant component(s)." ".Warnings, cautions and precautions: the implantation of spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.Proper patient selection is critical to the success of the procedure.Only patients who satisfy the criteria set forth under the indications section of this document and who do not have any of the conditions set forth under the contraindications section of this document should be considered for spinal fixation surgery using the nuvasive acp system.In addition, patients who smoke have been shown to have an increased incidence of pseudarthrosis.Based upon the fatigue testing results, the physician/surgeon should consider the levels of implantation, patient weight, patient activity level, other patient conditions, etc, which may impact the performance of the system." ".Follow proper instrument selection as specified in the surgical technique.The nuvasive acp system is designed to provide biomechanical stabilization as an adjunct to cervical fusion and should be used with anterior column support.Without anterior column support, its use may not be successful.Spinal fixation should only be undertaken after the surgeon has had hands on training in this method of spinal fixation and has become thoroughly knowledgeable about spinal anatomy and biomechanics.A surgical technique is available for instructions on the important aspects of this surgical procedure.Care should be taken to ensure that all components are ideally fixated prior to closure." ".Patient education: preoperative instructions to the patient are essential.The patient should be made aware of the limitations of the implant and potential risks of the surgery.The patient should be instructed to limit postoperative activity, as this will reduce the risk of bent, broken or loose implant components.The patient must be made aware that implant components may bend, break or loosen even though restrictions in activity are followed." ".Pre-operative warnings: only patients that meet the criteria described in the indications should be selected.Patient condition and/or predispositions such as those addressed in the aforementioned contraindications should be avoided.Care should be used in the handling and storage of the acp implants.The implants should not be scratched or damaged.Implants and instruments should be protected during storage and from corrosive environments.For sterile implants: assure highly aseptic surgical conditions, and use aseptic technique when removing the acp implant from its packaging.Inspect the implant and packaging for signs of damage, including scratched or damaged devices or damage to the sterile barrier.Do not use the acp implants if there is any evidence of damage.Care should be used during surgical procedures to prevent damage to the device(s) and injury to the patient." ".Post-operative warnings: damage to the weight-bearing structures can give rise to loosening of the components, dislocation and migration, as well as to other complications.To ensure the earliest possible detection of such catalysts of device dysfunction, the devices must be checked periodically postoperatively, using appropriate radiographic techniques." if any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NUVASIVE ACP SYSTEM
Common Device Name
APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY
Manufacturer (Section D)
NUVASIVE INC
7475 lusk blvd
san diego CA 92121
Manufacturer (Section G)
NUVASIVE INC
7475 lusk blvd
san diego CA 92121
Manufacturer Contact
geoff gannon
7475 lusk blvd
san diego, CA 92121
MDR Report Key18422677
Report Number2031966-2023-00294
Device Sequence Number18104519
Product Code KWQ
UDI-Device Identifier00195377009098
UDI-Public195377009098
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K201078
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/14/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number18008515
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/11/2023
Supplement Date Received by Manufacturer02/14/2024
Initial Report FDA Received Date12/29/2023
Supplement Report FDA Received Date02/14/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNKNOWN ACP PLATE
Outcome Attributed to Adverse Event Required Intervention;
Patient SexPrefer Not To Disclose
-
-