|
The device was not returned to olympus, and it is not expected to be returned for evaluation of the reported issue.Attempts to retrieve additional information from the customer are in progress.The investigation is ongoing, and a supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.The device was attached to the usg-400/esg-400 and a probe check was performed; the device failed the probe check.Both switches were checked and found to be functional.A visual inspection on the received condition was performed on the device; there is some tissue build up.The ptfe pad (teflon pad) was inspected and found to have normal wear with no metal exposed.The distal end of the device was inspected under the microscope, foreign material indicating the device had been used.The distal end of the device was inspected, and the probe was broken off.The missing portion was not returned.The wiper movement is normal.The consistency of movement of the handle and jaw is normal.The handle load is normal.The rotation of the knob torque is normal and smooth.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, it is likely that the probe broke due to contact with a surgical instrument or the non-insulated area of grasping section.Although a definitive root cause cannot be identified, the following step-by-step scenario likely caused the event: 1.The tissue pad was worn out because the seal and cut was output with the gripper closed (including after the tissue was cut) without grasping the tissue.2.Since the tissue pad was worn out, non-insulated area of the grasping section and the distal end of the probe came into contact.3.The output was activated in seal & cut mode in state of description stated above.Therefore, scratches (contact marks) were made on the probe and the grasping section, indicating that they came into contact with each other.4.The device was activated in seal &cut mode or while the grasping section was grasping tissue.This applied a force to the scratched area of the grasping section, causing this area to have cracks.5.A force was applied to the probe causing it to break.The event can be prevented by following the instructions for use (ifu) which state: ·do not activate output in seal & cut mode while the grasping section is closed without contacting tissue or vessel, or ensuring that tissue is transected.Otherwise, a local increase of the temperature due to a friction between the probe tip and the grasping section may result in various forms of damage in the probe tip and/or the tissue pad, such as premature wear, breakage, deformation, and/or falling off inside the body cavity and/or partial separating.¿ when cutting and vessel sealing is performed in seal & cut mode, apply light tension on the tissue so that users can confirm it is transected.Also, stop activation immediately after tissue is transected.Otherwise, the grasping section, the tissue pad, or the probe tip may break and fall off, and partial separating of the tissue pad may occur due to a local increase of temperature caused by the friction between tissue pad and the probe tip during activation.Olympus will continue to monitor field performance for this device.
|