|
MENTOR TEXAS MENTOR MEMORYSHAPE BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
|
Back to Search Results |
|
| Catalog Number |
354-1208G |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Deformity/ Disfigurement (2360)
|
| Date of Event |
09/28/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: breast prosthesis rupture.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Manufacturer¿s reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent a primary breast augmentation procedure with a mentor memoryshape breast implant 280cc gel breast prosthesis that ruptured post implantation.Rupture of the patient¿s right breast prosthesis was diagnosed during an office visit.As a result, the patient underwent explantation on (b)(6) 2023.
|
| |
|
Additional Manufacturer Narrative
|
|
On 16-jan-2024, the mentor failure analysis lab received the device for evaluation.On 18-jan-2024, mentor completed the investigation on the suspect medical device.According to the information received, the patient experienced a rupture in the breast implant.The product was returned to mentor for evaluation.Mentor conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that the breast implant was found to be ruptured.Additionally, shell wear was observed on the edges of the rupture.The evaluation determined that the possible cause of the rupture was consistent with normal wear.Shell wear suggests in-vivo folding or creasing of the device.This may be the result of the following factors: continuous and sustained stresses to the device such as too small breast pocket and folding or wrinkling of the shell in the breast pocket.In some cases, the breast implants may also wear out over time.A second product was received (lot#: 262870).No adverse events were reported for this concomitant (contralateral) device.Therefore, no further analysis is required.As part of mentor¿s quality process, all devices are manufactured, inspected, and released to approved specifications.No corrective and preventive action (capa) is required now.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of post-market surveillance.Manufacturer¿s reference number: (b)(4).
|
| |
|
Search Alerts/Recalls
|
|
|