• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC SOFAMOR DANEK USA, INC PIVOX¿ OBLIQUE LATERAL SPINAL SYSTEM; INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT,

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC SOFAMOR DANEK USA, INC PIVOX¿ OBLIQUE LATERAL SPINAL SYSTEM; INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, Back to Search Results
Model Number 2161001
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Device Embedded In Tissue or Plaque (3165)
Date of Event 12/07/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Information was received from healthcare provider (hcp) via a manufacturer representative regarding an event which occurred during u nknown spinal procedure.It was reported that this inserter tip sheared off when we implanted a pivox cage.The sheared off portion remains threaded into the cage that is implanted in the patient. the surgeon has no plans to revise because there is no way to reattach to the implant since the inserter tip is threaded into the implant holder. this instrument was not used during the surgery, because we looked and saw the instrument tip was chipped and didn't want to use it. there was no patient symptom reported.There were no further complications reported regarding the event.
 
Additional Manufacturer Narrative
H3: product analysis: part # 2161001, lot # nm18g004 visual and optical inspection confirmed the tip of the inserter has broken due to bend stress overload.H6: fdm and fdr was update annex g is updated medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PIVOX¿ OBLIQUE LATERAL SPINAL SYSTEM
Common Device Name
INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT,
Manufacturer (Section D)
MEDTRONIC SOFAMOR DANEK USA, INC
4340 swinea rd
memphis TN 38118
Manufacturer (Section G)
MEDTRONIC SOFAMOR DANEK USA, INC
4340 swinea rd
memphis TN 38118
Manufacturer Contact
glen belmer
1800 pyramid place
memphis, TN 38132
6122713209
MDR Report Key18424013
Report Number1030489-2024-00002
Device Sequence Number7771450
Product Code MAX
UDI-Device Identifier00643169639607
UDI-Public00643169639607
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K152277
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/11/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number2161001
Device Catalogue Number2161001
Device Lot NumberNM18G004
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/07/2023
Supplement Date Received by Manufacturer02/20/2024
Initial Report FDA Received Date01/02/2024
Supplement Report FDA Received Date03/11/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/20/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexFemale
-
-