• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZOLL MEDICAL CORPORATION X SERIES; DEFIBRILLATOR/PACEMAKER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL MEDICAL CORPORATION X SERIES; DEFIBRILLATOR/PACEMAKER Back to Search Results
Model Number X SERIES
Medical Device Problem Code No Display/Image (1183)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
 
Event or Problem Description
Complainant alleged that during a routine shift check by a clinician, the device's screen turned completely white and was illegible.Complainant indicated that there was no patient involvement in the reported malfunction.
 
Additional Manufacturer Narrative
The device was returned to zoll medical corporation for evaluation.The customer's report was not replicated or confirmed.The device was put through extensive testing including bench handling, impedance testing, defibrillation cycling, and environmental chamber testing without duplicating the report.An internal inspection of the device found no discrepancies.The device log does not capture reports of this nature.The lcd display was replaced as a precaution.The device was recertified and returned to the customer.Analysis of reports of this type has not identified an increase in trend.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
X SERIES
Common Device Name
DEFIBRILLATOR/PACEMAKER
Manufacturer (Section D)
ZOLL MEDICAL CORPORATION
269 mill road
chelmsford MA 01824
Manufacturer Contact
scott august
269 mill road
chelmsford, MA 01824
9784219552
MDR Report Key18424860
Report Number1220908-2023-05000
Device Sequence Number8014546
Product Code MKJ
UDI-Device Identifier00847946029500
UDI-Public00847946029500
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
P160022
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberX SERIES
Device Catalogue NumberX SERIES
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/15/2023
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/11/2023
Supplement Date Received by Manufacturer12/11/2023
Initial Report FDA Received Date01/02/2024
Supplement Report FDA Received Date01/18/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/01/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1