MENTOR TEXAS MENTOR SILTEX ROUND MODERATE PROFILE; PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE
|
Back to Search Results |
|
| Catalog Number |
3542680 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Deformity/ Disfigurement (2360)
|
| Date of Event |
12/06/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
At the time of this report, mentor has received no information regarding explantation or an expected explantation date.It is unknown at this time if the device will be made available for return.As a result, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation is in progress.Once completed, a supplemental report will be submitted.Reason for device explant and/or reoperation: as of now there is no information regarding the explant or reoperation.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Manufacturer¿s reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a 56-year-old female patient underwent primary breast augmentation with 475cc mentor siltex round moderate profile and experienced breast implant deflation on her left side postoperatively; diagnosed via photos.The patient has consulted with the healthcare professional.At the time of this report, mentor has received no information regarding explantation or an expected explantation date.
|
| |
|
Additional Manufacturer Narrative
|
|
On january 15, 2024, mentor became aware that the date of replacement surgery is (b)(6) 2024.Replacement order was placed for catalog number 3502650.The following fields have been updated on this form: - is required intervention has been selected under section b; field b2.- fields d6b for explantation date is (b)(6) 2024.- field h6 health effect - impact code ¿device explantation¿ has been added.- field h6 type of investigation has been updated to "device not returned".Reason for device explant and/or reoperation: left deflation.Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
On february 9, 2024, the mentor failure analysis lab received the device, and evaluation was completed as follows: mentor conducted a visual inspection, leak testing and microscopic examination of the returned device.Visual analysis of the returned sample revealed that no damage or anomalies were observed on the sal siltex rnd diap pkg 475cc breast implant.Leak testing was performed, according with mentor procedure, and it revealed a tear measuring approximately 0.7 cm on posterior view.Microscopic examination was performed, and the cause of the tear could not be identified.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.Although no conclusion could be reach on the cause of the reported event, the instructions for use contain the following caution: causes of deflation of saline implants include but are not limited to the following events: crease fold failure, intraoperative or postoperative trauma, excessive stresses or manipulations as may occur during normal, daily routines including customary and purposeful trauma such as that which can occur during vigorous exercise, athletics, etc.As part of mentor¿s quality process, all devices are manufactured, inspected, and released to approved specifications.No corrective and preventive action (capa) is required now.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post market surveillance.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Manufacturer¿s reference number: (b)(4).
|
| |
|