Please note corrections to section d4 (catalog#, lot#, gtin and product description).The reported event could be confirmed, since the conical extractor is broken as reported.The device inspection revealed the following: the visual inspection of the conical extractor has shown that the tip is broken off, the fragment was not returned for evaluation.The fracture surface runs obliquely, which is an indication of high lateral stress.The microscopic evaluation of the fracture face shows that the view is typical for a brittle overload fracture with a shear lip on one side.The shear lip also indicates that the device was exposed to high lateral forces when it broke.Based on that it can be concluded that a combination of high torsional force and lateral stress did lead to a mechanical overload and finally the breakage.Furthermore, a hardness test according to rockwell on the returned item indicates the material being in accordance to the values in the material specification.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.Based on investigation, the root cause was attributed to a user related issue.The event was ultimately caused by using a incorrect screwdriver for the removal of a by strongly tightened end cap that was implanted for more than one year.The initially used spreading recon lag screwdriver bit is cannulated and slotted at the hexagon and no designed to take such high forces.The slot in the hexagon had the effect that the tip was pushed together and as a result the force was no longer distributed as intended over the surfaces of the hexagon, which resulted in damage to the screwdriver and certainly also to the end cap recess.As stated in the t2 alpha femur antegrade operative technique is the set screw or end cap removed with the solid screwdriver bit, long, which is catalog # 2351-0110.After the end cap recess was damaged a removal with the standard screwdriver was not possible anymore.Therefore a end cap removal attempt with the compression screwdriver was made, but the hexagon of this screwdriver is also not designed for end cap removal, the tip smaller and much softer and get therefore easily damaged when excessive forces are applied.The conical extractor, which was used because of the damaged end cap recess, did break as too much torsional and lateral force was applied onto the device.If more information is provided, the case will be reassessed.
|