| Model Number |
9763 |
| Medical Device Problem Code |
Missing Test Results (3267)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/07/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The patient/family (a1) was the initial reporter (e1).So, personal information was not entered.No information was captured in section a4, as the customer's weight was not provided.
|
| |
|
Event or Problem Description
|
|
An advocate called on behalf of the customer, to report that one of their blood glucose readings was not stored in the memory of the contour next one meter.There was no allegation of an adverse event.The advocate was advised to return the device for evaluation.A replacement meter kit was sent to the customer.
|
| |
|
Additional Manufacturer Narrative
|
|
The customer returned the suspected contour next one meter with serial # (b)(6) for evaluation.No test strips were returned.Therefore, an in-house testing was performed with the returned meter and in-house contour next test strips using blood sample, which gave a satisfactory performance.The blood glucose readings obtained with the in-house testing were properly stored in the meter's memory.
|
| |
|