Zimvie complaint number (b)(4).A4: patient weight unknown / not provided.D6: explant date unknown / not provided.D10: concomitant medical product and therapy dates: 2 x int410, osseotite tapered certain implant 4 x 10mm, lot number: 2022080173; int3211, osseotite tapered certain implant 3.25 x 11.5mm, lot number: 2022080259.E1: reporter phone: (b)(6).H6: event problem codes - patient code 1932: inflammation.A summary investigation has been completed for peri-implantitis events recognizing that a definitive root cause cannot be identified due to a wide range of external factors (non-design or manufacturing related), including medical conditions (e.G., diabetes, poor bone quality, etc.) / patient habits (e.G., smoking) and surgical technique.Previously completed investigations for these events have not identified any signals indicating potential non-conformances affecting the manufacturing and sterilization processes.Furthermore, the probability of a manufacturing or design defect that might lead to peri-implantitis occurring and escaping the available detections has been assessed and found remote and almost nonexistent.Should additional information be received which indicates that the device may have caused or contributed to the event, an additional report will be submitted.H3 other text : summary investigation.
|
|
It was reported that the implants at tooth sites 13, 18, 21, 23 were removed due to peri-implantitis.Inflammation and edema were reported as a result of the event.Patient will have to return to place a new implant.
|