MAUDE Adverse Event Report: BIOMET 3I OSSEOTITE TAPERED CERTAIN IMPLANT 4 X 13MM; DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET 3I OSSEOTITE TAPERED CERTAIN IMPLANT 4 X 13MM; DENTAL IMPLANT Back to Search Results
Catalog Number INT413
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Unspecified Infection (1930); Osteolysis (2377); Swelling/ Edema (4577)
Date of Event 11/17/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Zimvie complaint number (b)(4).A4: patient weight unknown / not provided.D6: explant date unknown / not provided.D10: concomitant medical product and therapy dates: 2 x int410, osseotite tapered certain implant 4 x 10mm, lot number: 2022080173; int3211, osseotite tapered certain implant 3.25 x 11.5mm, lot number: 2022080259.E1: reporter phone: (b)(6).H6: event problem codes - patient code 1932: inflammation.A summary investigation has been completed for peri-implantitis events recognizing that a definitive root cause cannot be identified due to a wide range of external factors (non-design or manufacturing related), including medical conditions (e.G., diabetes, poor bone quality, etc.) / patient habits (e.G., smoking) and surgical technique.Previously completed investigations for these events have not identified any signals indicating potential non-conformances affecting the manufacturing and sterilization processes.Furthermore, the probability of a manufacturing or design defect that might lead to peri-implantitis occurring and escaping the available detections has been assessed and found remote and almost nonexistent.Should additional information be received which indicates that the device may have caused or contributed to the event, an additional report will be submitted.H3 other text : summary investigation.
 
Event or Problem Description
It was reported that the implants at tooth sites 13, 18, 21, 23 were removed due to peri-implantitis.Inflammation and edema were reported as a result of the event.Patient will have to return to place a new implant.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OSSEOTITE TAPERED CERTAIN IMPLANT 4 X 13MM
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
susanne taylor
4555 riverside drive
palm beach gardens, FL 33410
5619713230
MDR Report Key18426596
Report Number0001038806-2024-02549
Device Sequence Number8089211
Product Code DZE
UDI-Device Identifier00844868006664
UDI-Public(01)00844868006664(17)270601(10)2022051002
Combination Product (Y/N)N
Initial Reporter CountryPO
PMA/510(K) Number
K063341
Exemption Number5645646
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Dentist
Type of Report Initial
Report Date (Section B) 01/01/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Catalogue NumberINT413
Device Lot Number2022051002
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/12/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date01/02/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/31/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DENTAL IMPLANT
Patient Age35 YR
Patient SexFemale
Patient EthnicityNon Hispanic
-
-