• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TRIVIDIA HEALTH INC SYRINGES; PISTON SYRINGE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TRIVIDIA HEALTH INC SYRINGES; PISTON SYRINGE Back to Search Results
Model Number SYR, NPD 1.0CC 29G 100CT1/2" 8/CS
Medical Device Problem Code No Flow (2991)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/06/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Internal report reference number: (b)(4).Syringes were not returned for evaluation.Note: manufacturer contacted customer in a follow-up call on 22-dec-2023 to ensure the replacement products resolved the initial concern - able to establish contact with customer who stated replacement products resolved initial concern.
 
Event or Problem Description
Consumer reported complaint for the trueplus single-use insulin syringes.Customer stated the 29g syringes did not aspirate the insulin.The package was not open or damaged when received by the customer.The customer feels well and did not report any symptoms.The customer did not claim to be injured using the syringes and no medical attention associated with the use of the product was reported.
 
Additional Manufacturer Narrative
Sections with additional information as of; 25-mar-2024 h6: updated fda's type of investigation, investigation findings, and investigation conclusions.H10: complaint was forwarded to supplier quality based on complaint's description for investigations.Customer returned a used syringe; unable to ship to the manufacturer.Scrap returned product.Internal evaluation has been completed by the manufacturer.No abnormalities observed with retention.Added most likely underlying root cause onto case.Most likely underlying root cause: mlc-009: use error caused or contributed to event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYRINGES
Common Device Name
PISTON SYRINGE
Manufacturer (Section D)
TRIVIDIA HEALTH INC
2400 nw 55th court
fort lauderdale FL 33309
Manufacturer Contact
karen devincent
2400 nw 55th court
fort lauderdale, FL 33309
MDR Report Key18426890
Report Number1000113657-2024-00001
Device Sequence Number7743713
Product Code FMF
UDI-Device Identifier00021292005825
UDI-Public(01)00021292005825
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K993017
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Remedial Action Replace
Type of Report Initial,Followup
Report Date (Section B) 03/25/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberSYR, NPD 1.0CC 29G 100CT1/2" 8/CS
Device Lot NumberNP22333
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Distributor Facility Aware Date12/06/2023
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/06/2023
Supplement Date Received by Manufacturer03/06/2024
Initial Report FDA Received Date01/02/2024
Supplement Report FDA Received Date03/25/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-