INTEGRA LIFESCIENCES (IRELAND) LTD. EXCEL 23KHZ ANGLED HANDPIECE EACH1; ULTRASONIC SURGICAL PRODUCTS Back to Search Results
Catalog Number C2601
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/15/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
 
Event or Problem Description
This is 2 of 2 reports linked to mfg report number 3006697299-2023-00131: a facility reported that after the 1st sterilization, a crack was observed on the cusa excel 23khz angled handpiece (c2601) at the time of 1st surgery.The patient was already under anesthesia when the dysfunction was observed, so a backup handpiece was used.There was no patient injury; however, there was increased surgery time of 1 hour.
 
Additional Manufacturer Narrative
Updated fields: d4, d9, g3, g6, h2, h3, h4, h6, h10 the cusa handpiece (c2601) was not returned for evaluation as per customer; therefore, an evaluation of the device could not be performed.Lot number information has been provided; therefore, device history record (dhr) was reviewed, and no anomalies during the manufacture or packaging that could be associated with the complaint incident was observed.Based on the reported failure of crack on the handpiece, the complaint may have been a result of cracked housing which is pending corrective action.However, without testing device it is not possible to verify.If additional relevant information becomes available in the future, this complaint will be reopened, and the respective evaluation performed.Trends will be monitored for this and similar issues.At present, we consider this complaint to be closed.
 
Event or Problem Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXCEL 23KHZ ANGLED HANDPIECE EACH1
Common Device Name
ULTRASONIC SURGICAL PRODUCTS
Manufacturer (Section D)
INTEGRA LIFESCIENCES (IRELAND) LTD.
sragh tullamore, co. offaly
EI 
Manufacturer (Section G)
INTEGRA LIFESCIENCES (IRELAND) LTD.
sragh tullamore, co. offaly
EI  
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key18427362
Report Number3006697299-2023-00132
Device Sequence Number7856829
Product Code LFL
UDI-Device Identifier10381780039426
UDI-Public10381780039426
Combination Product (Y/N)N
PMA/510(K) Number
K141674
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberC2601
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/04/2023
Supplement Date Received by Manufacturer04/19/2024
Initial Report FDA Received Date01/02/2024
Supplement Report FDA Received Date04/26/2024
Date Device Manufactured08/25/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-