• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC MEXICO DUET®; DEVICE, MONITORING, INTRACRANIAL PRESSURE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC MEXICO DUET®; DEVICE, MONITORING, INTRACRANIAL PRESSURE Back to Search Results
Model Number 46913
Medical Device Problem Code Disconnection (1171)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/20/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received information regarding an external drainage and monitoring system.It was reported the patient was setup with the device and continuously monitored.At some point, the line by the stopcock broke where the tubing was separating from the piece it should be connected to.The device was replaced with a new one.There was a procedure delay of 15 minutes.The patient's status at the time of the report was alive-no injury.
 
Additional Manufacturer Narrative
H3.Product analysis determined that the visual inspection confirmed the tube had separated from the luer connector.There was adhesive present on the end of the luer medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DUET®
Common Device Name
DEVICE, MONITORING, INTRACRANIAL PRESSURE
Manufacturer (Section D)
MEDTRONIC MEXICO
av. paseo del cucapah #10510
tijuana,bc 22570
MX  22570
Manufacturer (Section G)
MEDTRONIC MEXICO
av. paseo del cucapah #10510
tijuana,bc 22570
MX   22570
Manufacturer Contact
justin ellis
9775 toledo way
irvine, CA 92618
7635265677
MDR Report Key18427484
Report Number9612164-2024-00014
Device Sequence Number15389444
Product Code GWM
UDI-Device Identifier00763000624767
UDI-Public00763000624767
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K802100
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/09/2025
Device Model Number46913
Device Catalogue Number46913
Device Lot Number224990849
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/26/2023
Supplement Date Received by Manufacturer08/23/2024
Initial Report FDA Received Date01/02/2024
Supplement Report FDA Received Date08/26/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/09/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexFemale
-
-