It was reported to boston scientific corporation that an autotome rx 49 was used in the papilla during a procedure on (b)(6) 2023.During the procedure, when the device was energized initially, the cutting wire of the autotome rx 49 broke.It was reported that no part of the cutting wire detached and fell into the patient.The procedure was completed with another of the same device.There were no patient complications reported as a result of this event.
|
|
Block e1 (initial reporter facility name): (b)(6).Block h6: imdrf device code a0401 captures the reportable event of cutting wire broken.Block h10: the returned autotome rx 49 was analyzed, and a visual evaluation noted that the cutting wire was broken blackened and kinked from the proximal pierce hole, which are consistent with the findings when the device was observed under magnification.Additionally, the working length was kinked, and the proximal pierce hole was torn.No other problems with the device were noted.The reported event of cutting wire break was confirmed.Upon analysis, it was found that the cutting wire was broken, kinked and blackened.Based on the condition of the device, the problem found could have been generated if there was contact between the device and the scope during energization or if the generator exceeded the maximum of voltage during the procedure.Also, energizing the device prior to performing sphincterotomy can compromise the cutting wire's integrity and cause a premature cutting wire fatigue.Once the cutting wire breaks, any attempt to remove the device from the scope can lead to the broken section hitting the working channel of the scope.This can bend the cutting wire.The working length was also found kinked, which could have been caused by operational factors, such as manipulating the device through difficult anatomy, the technique used during the device manipulation or by the interaction between the autotome and the scope (hitting against a hard surface).Additionally, the working length was torn at the proximal pierce hole, which could have been caused by submitting the cutting wire to tension during the handle actuation, or the possibility of the device being energized during the handle actuation.It is most likely that procedural or anatomical factors encountered during the use of the device could have affected the device performance and its integrity.Based on all gathered information, the most probable root cause of this complaint is adverse event related to procedure.A labeling review was performed and, from the information available, this device was used per the instructions for use (ifu) / product label.
|
|
It was reported to boston scientific corporation that an autotome rx 49 was used in the papilla during a procedure on (b)(6) 2023.During the procedure, when the device was energized initially, the cutting wire of the autotome rx 49 broke.It was reported that no part of the cutting wire detached and fell into the patient.The procedure was completed with another of the same device.There were no patient complications reported as a result of this event.
|