| Catalog Number |
71177733 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/07/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that, during an internal fixation an evos large tgtr 3.5mm linear hex drvr qc broke off in a large screwdriver handle while tightening a screw.The procedure was resumed, without any delay, using a s+n back-up device.It is unknown if pieces fell into the patient, but no injury was reported as a consequence of this issue.
|
| |
|
Additional Manufacturer Narrative
|
|
Internal complaint reference (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that, during an internal fixation an evos large tgtr 3.5mm linear hex drvr qc broke off in the handle while tightening a screw.The procedure was resumed, without any delay, using a s+n back-up device.No injury was reported as a consequence of this issue.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: additional information added in b5.H3, h6: although the involvement of all the devices was reported, the relationship with the investigated failure was directly associated to the evos hex driver.The device was not returned for evaluation; therefore, a device analysis could not be performed.The provided product identification information did not match any known release of this part and thus a manufacturing record review could not be conducted.A review of complaint history revealed similar events for the listed device over the previous 12 months.The batch number provided is not valid within the system, even though the review was performed and did not reveal similar events.This failure mode will be monitored for future complaints for any necessary corrective actions.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely potential factors that could contribute to the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|
| |
|
Search Alerts/Recalls
|