| Catalog Number |
5397190 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Bacteremia (4846)
|
| Date of Event |
08/24/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
H10: as the lot number for the device was provided, a review of the device history records will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.H10: d4 (expiry date: 09/2021).H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
|
| |
|
Event or Problem Description
|
|
It was reported through the results of a clinical that five months and six days post index procedure completion, the subject developed an adverse event of premature removal of tunneled central catheter due to suspected bacteremia.It was further reported that the patient was provided with antibiotics for treatment.Reportedly, the catheter was removed.The current status of the patient is unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
Manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: the physical device was not returned for evaluation.No photos were provided for review.Therefore, the investigation is inconclusive for the reported failure as no objective evidence was provided for review.Furthermore, the clinical conditions alleged in the complaint cannot be confirmed.The definitive root cause could not be determined based upon available information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.Section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported through the results of a clinical that five months and six days post an ultrasound-guided dialysis catheter placement via the right internal jugular vein, the subject developed an adverse event of premature removal of tunneled central catheter due to suspected bacteremia.It was further reported that the patient was provided with antibiotics for treatment.Reportedly, the catheter was removed.The current status of the patient is unknown.
|
| |
|